Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.
Recall
- Recall Number
- Z-1771-2026
- Event Number
- 98615
- Firm
- Boston Scientific Corporation
- FEI Number
- 2124215
- Product Code
- NKE
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- March 19, 2026
- Posted
- April 24, 2026
- Address
- 4100 Hamline Ave N, Saint Paul, MN, 55112-5700
Description
Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.
Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
Boston Scientific issued an Urgent Medical Device Correction notice to physicians, hospitals, and patients on 3/19/2026 via Federal Express or USPS certified mail. The notice explained the availability of the updated software, Brady Software maintenance release 6 (SMR6) for the ACCOLADE family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps), clinical and device longevity of the update, and recommendations for recommendations on the upgrade. In a separate notice, patients were advised to keep and attend all scheduled medical and device check appointments and talk to their doctor about the best steps for managing their pacemaker system. To confirm if the pacemaker or CRT-P is affected, please visit this website: www.BostonScientific.com/lookup. The language, as well as the model and serial numbers of the device, are required for the lookup which can be found on your patient ID card. Our Product Performance Resource Center at www.bostonscientific.com/ppr, includes information on this topic, a device lookup tool, and instructions for returning explanted products. If you have additional questions, or would like to report on a clinical event, please contact your Boston Scientific representative or Technical Services.
worldwide
685,776 units