FDA Recall Terminated

Bariatric Wheelchair, 24" Wide and 19" Deep Seat, 600 lb. Weight Capacity; Manufactured in Taiwan for Medline Industries, Inc., Mundelein, IL 60060 USA; Item #MDS809600 comes with Removable Desk-Length Arms, Swing-Away Detachable Footrests, Fixed Height. item #MDS809650 comes with Removable Desk-Length Arms, Swing-Away Detachable Elevating Legrests, Fixed Height.

Recall: Z-1770-2008 · Initiated March 31, 2008

Recall

Recall Number
Z-1770-2008
Event Number
47781
Firm
Medline Industries Inc
FEI Number
1417592
Product Code
IOR
Status
Terminated
Root Cause
Component design/selection
Initiated
March 31, 2008
Posted
August 19, 2008
Terminated
May 11, 2010
Address
1 Medline Pl, Mundelein, IL, 60060-4486

Description

Bariatric Wheelchair, 24" Wide and 19" Deep Seat, 600 lb. Weight Capacity; Manufactured in Taiwan for Medline Industries, Inc., Mundelein, IL 60060 USA; Item #MDS809600 comes with Removable Desk-Length Arms, Swing-Away Detachable Footrests, Fixed Height. item #MDS809650 comes with Removable Desk-Length Arms, Swing-Away Detachable Elevating Legrests, Fixed Height.

Reason

Fingers can get injured. There is a potential for users to pinch their fingers between the seat bars when opening the wheelchairs. Injuries can be severe from fractures to severing.

Action

A Voluntary Field Correction letter dated 3/31/08 was issued to customers. The letter informed them of reports of users catching their fingers under the seat bars, which are identified on the chair with warning stickers. The customers were requested to change the backrest with the enclosed replacement upholstery kit, properly dispose of the old back rest, and place the enclosed cautionary tags on both sides of the chair seat. Instructions for installation of the upholstery and cautionary tags were included, as was a sheet entitled "How to Open a Wheelchair" which gives step by step instructions with illustrations for opening the wheelchair, as well as the warning "CAUTION KEEP YOUR FINGERS ON TOP OF THE UPHOLSTERY AT ALL TIMES. DO NOT PLACE YOUR FINGERS ON THE SIDES OR BOTTOM OF THE CHAIR SEAT." The accounts were requested to complete the enclosed response sheet and fax it back to Medline at 847-643-4482, indicating the serial numbers of chairs corrected. Any questions were directed to 800-633-5463, extension 5349, between 8:00 AM and 5:00 PM CST. Distributors were requested to notify their accounts of the recall.

Distribution

Class II Recall - Worldwide Distribution --- including USA and Canada.

Quantity

430 wheelchairs