Siemens Medical Solutions, Mevatron KD2 Digital Linear Accelerator, part number 1940753 equipped with a Digital Electron Variable Applicator (DEVA, Part Number 8485971) with available energy of 6 MeV.
Recall
- Recall Number
- Z-1757-2008
- Event Number
- 47226
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2937457
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- December 10, 2007
- Posted
- September 29, 2008
- Terminated
- May 13, 2011
- Address
- 4040 Nelson Ave, Concord, CA, 94520-1200
Description
Siemens Medical Solutions, Mevatron KD2 Digital Linear Accelerator, part number 1940753 equipped with a Digital Electron Variable Applicator (DEVA, Part Number 8485971) with available energy of 6 MeV.
Radiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
Siemens sent a Customer Safety Advisory Notice dated 12/12/2007 to DEVA customers. The letter stated that Siemens would temporarily disable energies 5MeV and below for all customers that decide to continue using the DEVA, resulting in the temporary unavailability of energies 5MeV and below for all available. Customers wishing not to have these energies blocked must return their DEVA and they would be contacted by a local Siemens rep to schedule a service appointment. The letter stated that as a preventative measure that those customers should refrain from using the DEVA with any combination of energies 5MeV and below with field sizes smaller than 6cm x 6cm. The firm suggested using the fixed field electron applicators and cutouts with these energies and field sizes. Siemens will release a new control console software in 2008 for free-of-charge, that will limit the use of the DEVA to energies 6MeV and above, while allowing the use of lower energies using the fixed field electron applicatiors.
Worldwide Distribution: USA, Canada, Belgium, Bulgaria, France, Germany, Hungary, India, Italy, Japan, Malaysia, P.R. China, Pakistan, Poland, Portugal, Qatar, Republic Korea, South Africa, Spain, Sweden, and United Kingdom.
168 total