FDA Recall Terminated

KT RECOVERY+ ICE/HEAT MASSAGE BALL Rolling ball with hot and cold packs that can be interchanged depending on the therapy type desired

Recall: Z-1754-2020 · Initiated September 16, 2019

Recall

Recall Number
Z-1754-2020
Event Number
85315
Firm
KT Health, LLC
FEI Number
3007282994
Product Code
IME
Status
Terminated
Root Cause
Unknown/Undetermined by firm
Initiated
September 16, 2019
Terminated
October 1, 2021
Address
584 E 1100 S, Ste 4, American Fork, UT, 84003-3887

Description

KT RECOVERY+ ICE/HEAT MASSAGE BALL Rolling ball with hot and cold packs that can be interchanged depending on the therapy type desired

Reason

If over-heated may leak which, if not noticed, may cause hot fluid to come in contact with the user.

Action

Recall notifications were sent out on 9/16/2019 by letter to distributors. Firm indicated they used first class mail, email or facsimile. The notice informs users that the Company has received reports of product failures resulting in the release of the hot liquid from the Heat Gel-Core Insert component (orange ball) after being heated in the microwave. These included five reports of skin contact resulting in temporary or medically reversible adverse health consequences (first-degree burns and possibly one second degree burn), one of which required medical treatment. There were also four reports of skin contact with hot liquid that did not result in burns, and several reports of skin contact at room temperature with no adverse consequences The firm is asking Distributors to: 1) Immediately examine your inventory (in all locations), remove from sale/distribution and quarantine product subject to recall. 2) This recall should be carried out to the retailer and consumer level, so to the extent you have further distributed this product, please identify and alert your retailers and notify them at once of this product recall. Your notification to your retailers may be enhanced by including the enclosed sub-recall letter for them. 3) Please return all quarantined/recalled product to our warehouse at the following address for a full refund: KT Health, LLC Attn: Recall Returns 584 East 1100 South, Suite #4 American Fork, UT 84003 4) The firm will cover return shipping costs. If you have other costs kindly call us at: 801- 224-2717 5) Please acknowledge receipt of this recall letter by filling out the attached Return Response form and sending it by any of the following means: a. FAX: 858-947-3832 b. EMAIL: [email protected] c. MAIL: KT Health, LLC Attn: Recall Returns 584 East 1100 South, Suite #4 American Fork, UT 84003 All returns of full products will have the hot gel core insert balls removed and sent for destruction as noted above.

Distribution

US Nationwide distribution, no Government Consignees.

Quantity

94,554