FDA Recall Terminated

Prestilix 1600X Prestilix system is a fully integrated remote controlled digital radiology system. It is designed for all general radiology applications with or without contract injection and interventional procedures.

Recall: Z-1753-2019 · Initiated March 29, 2019

Recall

Recall Number
Z-1753-2019
Event Number
82668
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
KXJ
Status
Terminated
Root Cause
Device Design
Initiated
March 29, 2019
Posted
June 10, 2019
Terminated
April 15, 2021
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Prestilix 1600X Prestilix system is a fully integrated remote controlled digital radiology system. It is designed for all general radiology applications with or without contract injection and interventional procedures.

Reason

There has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.

Action

On March 29, 2019, GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated March 29, 2019 to its consignees/customers. The letter described the product, problem and actions to be taken. Customers were instructed to do the following: Please follow the precautions below until the monitor feet on your unit are replaced: -The CRT monitor and suspension should be inspected and used with caution. Make sure the monitors are secured properly to the tray before each use. -Immediately stop using the system if any looseness of the monitor is noticed and contact your GE service representative. -Monitor should not be positioned directly above the patient. -Before service or preventative maintenance of the monitor is attempted, review the service and preventive maintenance procedures. Also the customer were instructed to complete and return the MEDICAL DEVICE NOTIFICATION ACKNOWLEDGEMENT RESPONSE REQUIRED form by scanning or taking a photo of the completed form e-mailing to: [email protected]. F/U visits by GE Healthcare will be made to replace the plastic feet. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide Distribution: US (nationwide) to states of: AL, CA, CT, FL, GA, HI, IL, LA, MD, MI, MN, MO, NH, NY, OK, PA, TN, and VA; and countries of: Algeria, Argentina, Belgium, Bosnia & Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Cote D'Ivoire, Cyprus, Czech Republic, Egypt , France, French Polynesia, Germany, Greece, Guadeloupe, Guyana, Hungary, Italy, Japan, Korea, Lebanon, Mexico, Morocco, New Caledonia, Norway , Poland, Portugal, Russia, Saudi Arabia,Slovakia, Spain, Taiwan, Tunisia, Turkey, Ukraine, United Kingdom, Venezuela, and Vietnam.

Quantity

825 (38 US and 787 OUS) in total