FDA Recall Terminated

Hydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used in radiologic imaging procedures.

Recall: Z-1753-2013 · Initiated April 25, 2013

Recall

Recall Number
Z-1753-2013
Event Number
65176
Firm
Leibel-Flarsheim Company LLC
FEI Number
1518293
Product Code
OWB
Status
Terminated
Root Cause
Employee error
Initiated
April 25, 2013
Posted
July 18, 2013
Terminated
July 26, 2013
Address
2111 E Galbraith Rd, Cincinnati, OH, 45237-1624

Description

Hydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used in radiologic imaging procedures.

Reason

A complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum Fluoro Dose was found to be 30 R/min which is three times the Maximum Fluoro Dose limit of 10 R/min.

Action

The customers received an "Urgent Device Correction" notification letter on May 12, 2013 explaining the product issue and providing recommended actions. Customers were contacted to set up a service call to test and recalibrate equipment as necessary. Questions were directed to 800-778-7898.

Distribution

Nationwide Distribution (US), including the states of TN, KY, CO, KS, CT, HI, and AZ.

Quantity

12 units