Opiate 300 Ng Reagent (OP 1 X 250) for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems OP 300 Urine Calibrators, Catalogue No.475024. The OP assay provides a rapid screening procedure for determining the presence of OP and its metabolites in urine.

Recall: Z-1752-2020 · Initiated September 4, 2019

Recall

Recall Number
Z-1752-2020
Event Number
85289
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
DJG
Status
Terminated
Root Cause
Material/Component Contamination
Initiated
September 4, 2019
Posted
April 20, 2020
Terminated
April 28, 2021
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Opiate 300 Ng Reagent (OP 1 X 250) for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems OP 300 Urine Calibrators, Catalogue No.475024. The OP assay provides a rapid screening procedure for determining the presence of OP and its metabolites in urine.

Reason

Cross-reactivity testing may not meet the claims indicated in the Opiate 300 ng (OP)Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.

Action

An Urgent Medical Device Recall Letter was sent to customers via either U.S. postal service or email beginning on 9-4-2019. The letter informs customers that studies indicate that cross-reactivity testing may not meet the claims indicated in the Opiate 300 ng (OP) and Opiate 2000 ng (OP2) Instructions For Use (IFU). This letter included the Instructions for Use (IFU) claims versus study results for each product. Customers were informed of which compounds (at specific concentrations or ranges) may result in false positive or false negative test results. The letter indicated that Laboratories should be aware of these findings but that retrospective review of results is not required. Customers were informed that resolution of this issue was handled by revising the Opiate 300 ng (OP) Reagent IFU and the Opiate 2000 ng (OP2) Reagent IFU to update the Cross Reactivity sections. Customers were also requested to provide a copy of the letter to any other laboratory to which they may have forwarded affected product. Customers were requested to acknowledge receipt of the recall notification letter by responding to Beckman Coulter within 10 days either electronically, if the letter was received via email, or manually by completing and return the enclosed Response Form. Customers were instructed to contact the Beckman Coulter Customer Support Center through their website at http://www.beckmancoulter.com/customersupport/support; or by phone at 1-800-854-3633 in the United States and Canada with any questions. Customers outside of the U.S. and Canada were instructed to contact their local Beckman Coulter Representative.

Distribution

Distributed throughout the United States and to Puerto Rico. States include: CA, NC, OR, MI, GE, AL, SC, TX, NJ, IA, MS, IN, MA, NC, OH, AZ, LA, MA, NY, PA, KY, WV, MI, WA, CO, IL, VA, FL, KS, MD, ID, DE, NV, OK, AR, HI, WI, NY.

Quantity

7,951 units