Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56

Recall: Z-1751-2025 · Initiated April 11, 2025

Recall

Recall Number
Z-1751-2025
Event Number
96716
Firm
LeMaitre Vascular, Inc.
FEI Number
1220948
Product Code
DYG
Status
Open, Classified
Root Cause
Packaging process control
Initiated
April 11, 2025
Posted
May 12, 2025
Address
32 3rd Ave, Burlington, MA, 01803-4414

Description

Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56

Reason

During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier).

Action

LeMaitre notified OUS consignees beginning on 04/11/2025, and US consignees on 04/23/2025 via email and letter. Consignees were instructed to not use the devices, immediately quarantine any on hand, and complete and return the response form. Once received a LeMaitre Vascular representative will contact customers to provide return instructions. If the products were transferred or further distributed, customers were instructed to notify those customers and provide a copy of the recall letter.

Distribution

Nationwide distribution. International distribution to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, BULGARIA, CAMBODIA, CANADA, CHILE, COLOMBIA, COSTA RICA, CYPRUS, CZECHIA, DENMARK, DOMINICAN REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, HUNGARY, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA, MEXICO, MONGOLIA, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORTH MACEDONIA, NORWAY, OMAN, PANAMA, PARAGUAY, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKIYE, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.

Quantity

65,977 units