FDA Recall Terminated

HEMO-10 Hemospray Endoscopic Hemostat

Recall: Z-1746-2020 · Initiated February 4, 2020

Recall

Recall Number
Z-1746-2020
Event Number
84822
Firm
Wilson-Cook Medical Inc.
FEI Number
1037905
Product Code
QAU
Status
Terminated
Root Cause
Device Design
Initiated
February 4, 2020
Terminated
January 15, 2021
Address
4900 Bethania Station Rd, Winston Salem, NC, 27105-1203

Description

HEMO-10 Hemospray Endoscopic Hemostat

Reason

Complaints that the handle and/or activation knob on Hemostat devices have cracked or broken when the device is activated. In some cases, this has caused the carbon dioxide cartridge to exit the handle. Most of these complaints reported that the carbon dioxide cartridge exited the handle with minimal force. However, some complaints reported that the carbon dioxide cartridge exited the handle with the potential to project towards a user or a patient.

Action

Cook Medical notified customers initially via "URGENT: MEDICAL DEVICE RECALL" letters sent on about 02/04/2020 and 02/25/2020. The letters contained the same information and informed customers that Cook Medical has received complaints that the handle and/or activation knob on Hemospray Endoscopic Hemostat devices have cracked or broken when the device is activated. In some cases, this has caused the carbon dioxide cartridge to exit the handle. Instructions to customers included to examine inventory for affected devices, cease use of them and quarantine them, return affected devices immediately to Cook Medical, and complete and return the Acknowledgement and Receipt Form. Additionally customers were instructed to inform appropriate personnel and customers, if further distributed, of the recall down to the user level. Questions or concerns can be directed to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.

Distribution

Nationwide distribution. International distribution to Bahrain, Belarus, Canary Islands, Gibraltar, Guadeloupe, Guernsey, Iraq, Israel, Jersey, Jordan, Kuwait, Lebanon, Martinique, Monaco, Morocco, Oman, Pakistan, Qatar, Reunion, Saudi Arabia, Tunisia, United Arab Emirates, Unites States, Canada, Netherlands Antilles, Chile, Egypt, Japan, Mexico, Peru, Puerto Rico, Trinidad and Tobago, Australia, Brunei Dassalarum, Hong Kong, Korea, Myanmar, Macau, Malaysia, New Zealand, Philippines, Singapore, Thailand, Vietnam, Argentina, Brazil, Austria, Belgium, Bulgaria, Switzerland, Cyprus, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Hungary, Ireland, Iceland, Italy, Luxembourg, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Sweden, Slovenia, Slovakia, Turkey, Croatia

Quantity

89,423 devices total