INOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, Model Number: 10007
Recall
- Recall Number
- Z-1745-2022
- Event Number
- 90723
- Firm
- Mallinckrodt, LLC.
- FEI Number
- 3014074925
- Product Code
- MRN
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- June 23, 2022
- Posted
- September 13, 2022
- Address
- 1425 Us Highway 206, Bedminster, NJ, 07921-2653
Description
INOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, Model Number: 10007
The INOmax DSIR Delivery System, manufactured by Mallinckrodt Pharmaceuticals, is not compatible with the upgraded version of the Leoni Plus ventilator, manufactured by Lowenstein Medical. The Leoni Plus ventilator was recently upgraded in 2021 by Lowenstein Medical, with new hardware and software, and the upgraded version cannot be appropriately connected with the INOmax DSIR as instructed in the INOmax DSIR Operators Manual (OM). The INOmax DSIR OM requires the use of a one-way valve with the Leoni Plus ventilator during High Frequency Oscillatory Ventilation mode and warns that use of the device without a one-way valve may result in high Nitric Oxide (NO) delivery.
Mallinkrodt issued via email on 6/23/22 an Urgent Field Safety Notice (FA-R-0023) stating reason for recall, health risk and action to take: Upon receipt of this Field Safety Notice (FSN), INOmax DSIR operators are advised not to use an upgraded Leoni Plus ventilator with the INOmax DSIR. The INOmax DSIR Validated Ventilator List, supplied as Technical Bulletin TB-20005, will be revised to remove the Leoni Plus ventilator and distributed as soon as it is available. A response form is required for this FSN. Please complete the attached form and return as instructed.
Foreign Only: Argentina Australia Austria Beligum Canada Chile Columbia Czech Republic Denmark Ecuador Estonia Finland France Germany Great Britain Hungary Iceland Ireland Italy Latvia Mexico Netherlands Norway Poland Portugal Romania Slovakia South Africa Spain Sweden Switzerland UAE Uruguay
4,225 units OUS