Siemens Medical Solutions USA, Inc., ACUSON AcuNav ultrasound catheters when using SwiftLink catheter connectors; Model Number: 10348560.
Recall: Z-1745-2011
·
Initiated March 20, 2009
Recall
- Recall Number
- Z-1745-2011
- Event Number
- 52904
- Firm
- Siemens Medical Solutions USA, Inc.
- FEI Number
- 2936884
- Product Code
- DQO
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- March 20, 2009
- Posted
- March 23, 2011
- Terminated
- April 1, 2011
- Address
- 1230 Shorebird Way P.O. Box 7393, Mountain View, CA, 94043
Description
Siemens Medical Solutions USA, Inc., ACUSON AcuNav ultrasound catheters when using SwiftLink catheter connectors; Model Number: 10348560.
Reason
Screen Image Quality. An assembly issue on SwiftLink connectors may cause reduced/degraded image quality when using ACUSON AcuNav ultrasound catheter transducers.
Action
Siemens Field correction was initiated on 3/20/2009. A device correction letter was distributed to all affected users, with a description of the problem and its corrective actions. Siemens service personnel will make a site visit.
Distribution
Distribution in US and Internationally.
Quantity
21 units