FDA Recall Open, Classified

RayStation 8.0.0.61, 8.0.1.10. Radiation Therapy Treatment Planning System.

Recall: Z-1742-2024 · Initiated March 28, 2024

Recall

Recall Number
Z-1742-2024
Event Number
94388
Firm
RAYSEARCH LABORATORIES AB Eugeniavagen 18c Stockholm Sweden
FEI Number
3007774465
Product Code
MUJ
Status
Open, Classified
Root Cause
Software design
Initiated
March 28, 2024
Posted
May 1, 2024

Description

RayStation 8.0.0.61, 8.0.1.10. Radiation Therapy Treatment Planning System.

Reason

Potential for reported SSD to be too high.

Action

A Field Safety Notice for Medical Device Correction #133261, dated 3/28/24, was sent to consignees via email. Consignees were provided the instructions to use the Measure Tool to measure the distance in the patient view to make sure the reported SSD value is correct if it is to be used for patient setup or other critical tasks. If the SSD value is found to be incorrect, small changes to the isocenter, beam angles or the affected ROI's geometry can be made until the correct value is achieved. Consignees are to inform planning staff and all users of the workarounds provided in the Field Safety Notice. Consignees are to inspect their product and identify if installed units possess the affected software versions. Consignees are to reply to the recall notification confirming that they read and understood the Field Safety Notificaiton. RaySearch is to deploy a new version of RayStation/RayPlan to correct the issue, to be released in April 2024. The provided Field Safety Notice is to be maintained with affected units. Consignees with any questions are to contact [email protected].

Distribution

Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand, Norway, Poland, Qatar, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK.

Quantity

9 units