FDA Recall Terminated

ReNew Endocut Scissor Disposable Tip, Catalog Number: 3142, UDI: 00811099012115 The ReNew single patient use disposable tips are to be used with the ReNew Laparoscopic Hand Pieces and they are indicated for cutting, grasping, dissecting and coagulation of tissue in endoscopic and laparoscopic surgical procedures.

Recall: Z-1741-2018 · Initiated November 1, 2017

Recall

Recall Number
Z-1741-2018
Event Number
79435
Firm
Microline Surgical, Inc.
FEI Number
1000137656
Product Code
GEI
Status
Terminated
Root Cause
Process control
Initiated
November 1, 2017
Terminated
September 17, 2019
Address
50 Dunham Rd, Suite 1500, Beverly, MA, 01915-1882

Description

ReNew Endocut Scissor Disposable Tip, Catalog Number: 3142, UDI: 00811099012115 The ReNew single patient use disposable tips are to be used with the ReNew Laparoscopic Hand Pieces and they are indicated for cutting, grasping, dissecting and coagulation of tissue in endoscopic and laparoscopic surgical procedures.

Reason

Certain lots of disposable laparoscopic instrument tips may potentially have a compromised sterile barrier.

Action

Microline Surgical sent an Urgent Medical Device Recall letter dated November 1, 2017. Actions to be taken by the Customer/User: Please locate those products you have, quarantine them, and arrange for their return by contacting Microline Customer Service to receive an RMA number. Alternatively, you may wish to contact your Microline Distributor or Sales Representative to help with this. If you have used some or all of the identified product, please provide the item number, lot number, and the amount used so that we can track the progress of the recall. T To date, we have no evidence to suggest that the sterile barrier provided by the sealed pouches were damaged because of the rework, but are initiating this recall out of an abundance of caution because if a sterile barrier was compromised, there is potential that use of the device could result in an infection. If you experience any adverse events or quality problems with the use of this product, you should report this to FDAs MedWatch Adverse Event Reporting Program and to Microline as soon as possible. All product returned will be replaced at no cost. We regret any inconvenience and encourage you to contact Microline Customer Service at 978- 922-9810 or inf o@m icr olinesur g ical.com. with any further questions. For further questions, please call (978) 867-1758.

Distribution

Distributed in the following U.S. states: FL, MA, MI, OH, OK TN, and Netherlands.

Quantity

4 boxes of 10 tips each