FDA Recall Open, Classified

IntelliSpace Cardiovascular software versions 6.x, 7.x, and 8.x; Model Number: 830089.

Recall: Z-1740-2025 · Initiated April 7, 2025

Recall

Recall Number
Z-1740-2025
Event Number
96377
Firm
Philips Medical Systems Nederland B.V. High Tech Campus 36 Eindhoven Netherlands
FEI Number
3000976525
Product Code
LLZ
Status
Open, Classified
Root Cause
Software in the Use Environment
Initiated
April 7, 2025
Posted
May 8, 2025

Description

IntelliSpace Cardiovascular software versions 6.x, 7.x, and 8.x; Model Number: 830089.

Reason

An issue was identified with the software when using the Echo Module of ISCV, the issue may result in missing or incomplete information in the report.

Action

On April 7, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. 1. Actions that should be taken by the customer / user in order to prevent risks for patients or users " Verify that text added into the report conclusion matches the data in the work area interpret sheet (Refer to Attachment A Screenshot 2) " You may continue to use your system(s) in accordance with the intended use and by following the recommendation above. " Circulate this notice to all users of this device so that they are aware of the potential issue. " Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. " Please review the workflows described in Attachment A and identify whether either workflow is currently in use, or planned to be used, at your facility. Please note your response in the attached customer response form. " If these are not planned to be used at this time, please contact Philips Informatics Customer Support and ensure that this correction is implemented before beginning to use these workflows. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt of this letter via email to: [email protected]. 2. Actions planned by Philips to correct the problem Once all customer responses have been received, a Philips representative will contact customers who use or may use the affected workflows to schedule a time to install the software solution on your system(s) to resolve the issue (reference FCO-TBD). Please be assured that maintaining the highest level of safety and quality is our greatest priority. If you need any further information or support concerning this issue, please contact your local Philips representative: Philips Informatics Customer Support: 1-800-669-1328, Option 1. This notice has been reported to the appropriate Regulatory Agencies. Adverse reactions or quality problems experienced with the us

Distribution

Domestic: Nationwide Distribution International: Australia, Austria, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Indonesia, Ireland, Italy, Japan, Luxembourg, Mexico, Netherlands, Norway, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Quantity

996 units (267 US, 729 OUS)