FDA Recall
Terminated
Axiom Artis systems with Motor Controller Unit. Siemens Medical Solutions USA, Inc. Malvern, PA 19355. Intended for x-ray, angiographic use.
Recall: Z-1732-2010
·
Initiated April 12, 2010
Recall
- Recall Number
- Z-1732-2010
- Event Number
- 55647
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- MQB
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- April 12, 2010
- Posted
- May 28, 2010
- Terminated
- April 11, 2013
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355
Description
Axiom Artis systems with Motor Controller Unit. Siemens Medical Solutions USA, Inc. Malvern, PA 19355. Intended for x-ray, angiographic use.
Reason
sporadic malfunctions of the Artis Systems with Motor Controller caused by tolerances on electronic parts that might increase during the lifetime of the system.
Action
Siemens issued an "Urgent Field Safety Notice" to consignees describing the affected device instructing users to contact the recalling firm's service organization to schedule maintenance. For further information, contact Siemens Medical Solutions USA, Inc. at 1-610-448-3237.
Distribution
Nationwide Distribution
Quantity
83 units