Hemochron Jr. ACT- LR Assay (JACT- LR); ITC Catalogue/Code Number: JACTLR 45 Cuvettes; For Performance on the HEMOCHROM Whole Blood Microagulation Systems; International Technidyne Corp., Edison, NJ 08820. The HEMOCHRON Jr. ACT-LR is a quantitative assay for monitoring heparin anticoagulation during various medical procedures. The ACT-LR demonstrated linear correlation to the anticoagulation effects of heparin up to 2.5 unit/ml of blood. It is intended for use in monitoring low to moderate heparin doses frequently associated with procedures such as cardiac catheterization, Extracorporeal Membrane Oxygenation (ECMO) Hemodialysis, and Percutaneous Transluminal Coronary Angioplasty.

Recall: Z-1732-2009 · Initiated May 5, 2009

Recall

Recall Number
Z-1732-2009
Event Number
52018
Firm
International Technidyne Corp.
FEI Number
1000526865
Product Code
JBP
Status
Terminated
Root Cause
Employee error
Initiated
May 5, 2009
Posted
August 21, 2009
Terminated
September 30, 2009
Address
6-8 Olsen Avenue, Edison, NJ, 08820-2419

Description

Hemochron Jr. ACT- LR Assay (JACT- LR); ITC Catalogue/Code Number: JACTLR 45 Cuvettes; For Performance on the HEMOCHROM Whole Blood Microagulation Systems; International Technidyne Corp., Edison, NJ 08820. The HEMOCHRON Jr. ACT-LR is a quantitative assay for monitoring heparin anticoagulation during various medical procedures. The ACT-LR demonstrated linear correlation to the anticoagulation effects of heparin up to 2.5 unit/ml of blood. It is intended for use in monitoring low to moderate heparin doses frequently associated with procedures such as cardiac catheterization, Extracorporeal Membrane Oxygenation (ECMO) Hemodialysis, and Percutaneous Transluminal Coronary Angioplasty.

Reason

One lot of ACT-LR was released when one finished good testing criteria acceptance limit did not meet acceptance criteria.

Action

An "Urgent - Medical Device Recall" letter dated May 4, 2009 was sent to healthcare professionals. The recall letter described the reason for recall and the possible health risks involved. Consignees were asked to check their inventory for the affected lot number and return the completed Customer Account Tracking form via mail, e-mail or fax to ITC. ITC will make arrangements for the return of affected product. If there is no affected product in the inventory and no product or safety concerns, the letter is for information purposed only. Note: the Customer Account Tracking form should be completed in all cases. If you have questions regarding the recall information, please call ITC Technical Support at 1-800-579-2255, or email ITC at [email protected].

Distribution

Nationwide.

Quantity

173 boxes of 45 cuvettes