Maquet PowerLED 700. Product Usage; To provide illumination to the surgical area or the patient.
Recall
- Recall Number
- Z-1730-2014
- Event Number
- 68142
- Firm
- Maquet Medical Systems USA
- FEI Number
- 3008355164
- Product Code
- FSY
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 16, 2014
- Posted
- June 6, 2014
- Terminated
- April 6, 2016
- Address
- 45 Barbour Pond Dr, Wayne, NJ, 07470-2094
Description
Maquet PowerLED 700. Product Usage; To provide illumination to the surgical area or the patient.
Maquet has identified an issue with the Maquet PowerLED 700 (certain serial numbers). One or several plastic fixation tabs can break, which can cause the ambient light module to partially detach.
Maquet issued "Urgent-Medical Device Field Correction" notices, dated April 16, 2014 to affected customers via Federal Express. The notice identified the product, related issue and the action needed to be taken by the customer. The corrective action will include inspection of the fixation tabs of the Ambient Light Module adn replacement of the Ambient Light Module if tabs are found to be defective. The correction will be performed by an authorized MAQUET Service Technician. MAQUET Service Team representatives will schedule on-site service with customers. If customers have any questions or require additional information, the notice instructs them to contact their local MAQUET representative, or MAQUET Customer Service at 1-888-627-8383 (option 3), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST.
US Nationwide Distribution.
United States 1169 units