FDA Recall Terminated

Alere Triage D-Dimer Test PN 98100, Lot W53884B. The Alere Triage D-Dimer Test is a fluorescence immunoassay to be used with the Alere Triage Meters for the quantitative determination of cross-linked fibrin degradation products containing D-dimer in EDTA anticoagulated whole blood and plasma specimens. . The Alere Triage D-Dimer test is intended for use as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism.

Recall: Z-1727-2014 · Initiated May 2, 2014

Recall

Recall Number
Z-1727-2014
Event Number
68230
Firm
Alere San Diego, Inc.
FEI Number
1000125596
Product Code
GHH
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 2, 2014
Posted
June 6, 2014
Terminated
February 27, 2015
Address
9975 Summers Ridge Rd, San Diego, CA, 92121-2997

Description

Alere Triage D-Dimer Test PN 98100, Lot W53884B. The Alere Triage D-Dimer Test is a fluorescence immunoassay to be used with the Alere Triage Meters for the quantitative determination of cross-linked fibrin degradation products containing D-dimer in EDTA anticoagulated whole blood and plasma specimens. . The Alere Triage D-Dimer test is intended for use as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism.

Reason

Alere initiated this recall because a limited number of Alere Triage D-Dimer devices from PN 98100 Lot W53884B were incorrectly manufactured. As a result, when testing patient sample these affected devices will either generate an error code (e.g. E4 error), or they may generate an incorrect result.

Action

Alere sent an Urgent Medical Device Recall letter dated May 2, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are instructed to stop using all product from the affected lot, lot W53884B, and to discard any unused product. The customers were also instructed to complete and return the verification form indicating any credit or replacement product requested. They were also instructed to contact their primary consignees and advise them of the notice. Should you have any questions about the information contained in this notification, please call (877) 308-8287. and /or email:[email protected]

Distribution

Nationwide Distribution only.

Quantity

373 kits