FDA Recall Terminated

B-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitative determination of degradation products of type I collagen in human serum and plasma as an aid in assessing bone resorption. The test may be used as an aid in monitoring antiresorptive therapies (e.g. bisphosphonates, hormone replacement therapy - HRT) in postmenopausal women and individuals diagnosed with osteopenia. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.

Recall: Z-1727-2013 · Initiated April 25, 2013

Recall

Recall Number
Z-1727-2013
Event Number
65163
Firm
Roche Diagnostics Operations, Inc.
FEI Number
1823260
Product Code
JMM
Status
Terminated
Root Cause
Device Design
Initiated
April 25, 2013
Posted
July 12, 2013
Terminated
September 27, 2013
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

B-CrossLaps/serum (B-CTx in serum) REF 11972308 160 100 tests Immunoassay for the in vitro quantitative determination of degradation products of type I collagen in human serum and plasma as an aid in assessing bone resorption. The test may be used as an aid in monitoring antiresorptive therapies (e.g. bisphosphonates, hormone replacement therapy - HRT) in postmenopausal women and individuals diagnosed with osteopenia. The electrochemiluminescence immunoassay ECLIA is intended for use on Elecsys and cobas e immunoassay analyzers.

Reason

Roche Diagnostics internal investigations in R&D have shown that the claims for biotin interference and the sample stability given in the Elecsys -CrossLaps (-CL) assay package insert version 5, could not be reproduced. The -CrossLaps test results contribute to the monitoring of antiresorptive therapy. False low or false normal results may lead to delayed diagnosis and treatment of osteoporosi

Action

The firm, Roche, sent an "URGENT MEDICAL DEVICE CORRECTION" (UMDC)letter dated May 2, 2013, to its consignees/customers via UPS Ground. The letter describes the product, problem and actions to be taken. The consignees/customers were instructed to apply the correct sample storage conditions as stated in the UMDC and to inform all relevant parties about the correct biotin interference claim. The consignees/customers were instructed to keep the UMDC on file for reference until the revised package insert is available. The consignees/customers were asked to complete and return a form via fax (1-888-943-4205). Consignees/customers were instructed to call Roche Diagnostics Technical Support, 24 hours a day, seven days a week, at 1-800-428-2336 if they had any questions about the recall. Non-responding accounts are monitored on an ongoing basis.

Distribution

Nationwide distribution: USA including states of: AL, CA, GA, FL, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, TX, VA, VT, and WA.

Quantity

3056 units