FDA Recall Terminated

PhD System EIA/IFA Software; Model/Catalog Number: 426-0246; The Bio-Rad PhD System is an EIA and IFA processing system that consists of a PhD fluid handling station(s) linked to a network computer via an Ethernet hub. The network computer provides worklist generation, data management, data reduction and microplate reader control functions. Product is manufactured by Bio-Rad Laboratories, Inc., Espace European De L Enterprise Strasbourg Schiltigheim, FRANCE

Recall: Z-1726-2011 · Initiated January 20, 2011

Recall

Recall Number
Z-1726-2011
Event Number
57778
Firm
Bio-Rad Laboratories, Inc.
FEI Number
2915274
Product Code
JQW
Status
Terminated
Root Cause
Other
Initiated
January 20, 2011
Posted
March 18, 2011
Terminated
August 17, 2011
Address
4000 Alfred Nobel Dr, Hercules, CA, 94547-1803

Description

PhD System EIA/IFA Software; Model/Catalog Number: 426-0246; The Bio-Rad PhD System is an EIA and IFA processing system that consists of a PhD fluid handling station(s) linked to a network computer via an Ethernet hub. The network computer provides worklist generation, data management, data reduction and microplate reader control functions. Product is manufactured by Bio-Rad Laboratories, Inc., Espace European De L Enterprise Strasbourg Schiltigheim, FRANCE

Reason

Firm's Correction involves advising customers who have implemented a Laboratory Information System (LIS) that they need to ensure quantitative results have passed all validation criteria prior to reporting patient results. When a patient result fails a Validation Rule, the qualitative result will indicate "INVALID".

Action

The firm, Bio-Rad, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated January 20, 2011 to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to review the PhD Worklist Results to verify that all validations have passed and that the quantitative results are valid; to translate (if necessary) the customer letter, add local contacts information, distribute to all active PhD System customers in their region and provide tracking documentation to CSD-RA for all customers who received the customer letter. If you have any questions, contact the Bio-Rad CSD Regulatory Affairs Department at (510) 741-4618.

Distribution

Worldwide distribution: USA and countries including:Canada, China, Australia, New Zealand, Czech Republic, France, Hong Kong, India, Italy, Korea, Singapore, Sweden, Taiwan and United Kingdom.