FDA Recall Terminated

Hologic Explorer series Bone Densitometer System Software versions 12.6, 2.3., or 2.4. Indicated for the estimation of bone mineral density (BMD), comparison of measurable variables obtained from a given QDR scan to a database of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics.

Recall: Z-1726-2009 · Initiated April 7, 2009

Recall

Recall Number
Z-1726-2009
Event Number
51692
Firm
Hologic, Inc.
FEI Number
1221300
Product Code
KGI
Status
Terminated
Root Cause
Component design/selection
Initiated
April 7, 2009
Posted
August 25, 2009
Terminated
November 29, 2010
Address
35 Crosby Dr, Bedford, MA, 01730

Description

Hologic Explorer series Bone Densitometer System Software versions 12.6, 2.3., or 2.4. Indicated for the estimation of bone mineral density (BMD), comparison of measurable variables obtained from a given QDR scan to a database of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics.

Reason

System may not turn off at the end of the x-ray and cause additional x-ray exposure

Action

Hologic issued an Urgent: Medical Device Recall customer notification letter, dated April 17, 2009, by certified mail to inform customers of the problem. With the notification letter, customers will receive the following: -Software version Apex 12.6.2, 2.3.2, or 2.4.2 -Instructions to upgrade their systems with the provided software and to verify the new version on the system as a confirmation that the software was successfully loaded. -New Operator's Manual for installations where the operating characteristics of the system have changed due to the software upgrade. For Customers with systems with software versions other than 12.6, 2.3, or 2.4: Customers were asked to upgrade their QDR software, return their old QDR software disk, and complete and return to Hologic the enclosed Confirmation of Notification.

Distribution

Worldwide Distribution -- USA, Canada, Europe, Austria, Belgium, Bulgaria, Czech Republic, Denmark, France, Germany, Great Britain, Greece, Hungary, Iceland, Italy, Latvia, Lithuania, Macedonia, Netherlands, Norway, Poland, Portugal, Romania, Russia, Serbia, Slovak Republic, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine, Uzbekistan, Algeria, Arab Emirates, Argentina, Australia, Bahrain, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Egypt, Hong Kong, Indonesia, India, Iran, Iraq, Israel, Japan, Jordan, Kenya, Korea, Kuwait, Lebanon, Malaysia, Morocco, Mexico, Pakistan, Peru, Puerto Rico, South Africa, Saudia Arabia, Singapore, Syria, Taiwan, Thailand, Tunisia, Venezuela, and Vietnam.

Quantity

1914 units total