FDA Recall
Open, Classified
Canon Aquilion Prime, Model TSX-303A
Recall: Z-1723-2020
·
Initiated March 25, 2019
Recall
- Recall Number
- Z-1723-2020
- Event Number
- 84954
- Firm
- Canon Medical System, USA, INC.
- FEI Number
- 2020563
- Product Code
- JAK
- Status
- Open, Classified
- Root Cause
- Software design
- Initiated
- March 25, 2019
- Posted
- April 10, 2020
- Address
- 2441 Michelle Dr, Tustin, CA, 92780-7047
Description
Canon Aquilion Prime, Model TSX-303A
Reason
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
Action
Recall letters dated 3/25/2019 were issued on 3/25/2019 via USPS return receipt mail.
Distribution
Distribution was made to AK, AR, CA, FL, IA, ID, IL, KS, MA, MD, MI, MN, MO, MT, NC, ND, NH, NY, OH, OK, OR, PA, TN, TX, VT, WA, WI, and WV. There was government and military distribution but no foreign distribution.
Quantity
10 devices