Pericardiocentesis Catheter Tray, RPN C-PCYS-850, Order # G03284

Recall: Z-1723-2019 · Initiated April 9, 2019

Recall

Recall Number
Z-1723-2019
Event Number
82692
Firm
Cook Inc.
FEI Number
1820334
Product Code
PYC
Status
Terminated
Initiated
April 9, 2019
Terminated
April 28, 2020
Address
750 N Daniels Way, Bloomington, IN, 47404-9120

Description

Pericardiocentesis Catheter Tray, RPN C-PCYS-850, Order # G03284

Reason

Affected products may have been manufactured with the catheter distal end hole too small

Action

On April 9, 2019 Cook Medical issued URGENT MEDICAL DEVICE RECALL notices to customers via courier service. Customers were advised to take the following Actions: 1. Examine your inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of these products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. NOTE: Unaffected products that are returned will not be credited. 3. Complete and return the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return it via fax (812.339.7316) or by email to [email protected]. 4. The notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time) or by email to [email protected].

Distribution

Domestically: OH; Internationally: CANADA, SPAIN, ITALY, BELGIUM, GERMANY, UNITED ARAB EMIRATES, GREECE, SWITZERLAND, OMAN, MONACO, SAUDI ARABIA, AUSTRALIA, HONG KONG, INDIA & KOREA

Quantity

2 kits