FDA Recall Terminated

The LIFEPAK 20 defibrillator/monitor and LIFEPAK 20e. Product part numbers: 3202487-xxx, 3202488-000, U3202487-000, 3202488-015, and U3202488-000. The LIFEPAK 20 defibrillator/monitor and LIFEPAK 20e are acute cardiac care devices used by authorized healthcare providers in hospital and clinic settings. It is commonly found on "code" or "crash" cars in hospital hallways, plugged into AC power. In an emergency the user retrieves the cart, unplugging the monitor, and uses the device with the door closed for easy 1-2-3 operation. When the code team arrives the user will open the door and have access to all functions, including ECG monitoring. In addition to functioning as a semi-automatic defibrillator the LIFEPAK 20/20e provides options for: ECG monitoring, manual mode defibrillation; Non-invasive pacing; synchronized cardioversion; pulse oximetry (SpO2); hard paddles (adult and pediatric) and internal paddle accessories; printing; and backup batter power. The units are biphasic waveform devices

Recall: Z-1722-2011 · Initiated May 27, 2010

Recall

Recall Number
Z-1722-2011
Event Number
55875
Firm
Physio Control, Inc.
FEI Number
3015876
Product Code
MKJ
Status
Terminated
Root Cause
Device Design
Initiated
May 27, 2010
Posted
March 18, 2011
Terminated
September 29, 2014
Address
11811 Willows Rd Ne, Redmond, WA, 98052-2003

Description

The LIFEPAK 20 defibrillator/monitor and LIFEPAK 20e. Product part numbers: 3202487-xxx, 3202488-000, U3202487-000, 3202488-015, and U3202488-000. The LIFEPAK 20 defibrillator/monitor and LIFEPAK 20e are acute cardiac care devices used by authorized healthcare providers in hospital and clinic settings. It is commonly found on "code" or "crash" cars in hospital hallways, plugged into AC power. In an emergency the user retrieves the cart, unplugging the monitor, and uses the device with the door closed for easy 1-2-3 operation. When the code team arrives the user will open the door and have access to all functions, including ECG monitoring. In addition to functioning as a semi-automatic defibrillator the LIFEPAK 20/20e provides options for: ECG monitoring, manual mode defibrillation; Non-invasive pacing; synchronized cardioversion; pulse oximetry (SpO2); hard paddles (adult and pediatric) and internal paddle accessories; printing; and backup batter power. The units are biphasic waveform devices

Reason

A component failure on the AC power supply assembly results in "NO AC POWER" operation. A failure of AC power can result in a delay of defibrillation therapy if backup battery power is depleted.

Action

PhysioControl, Inc. began sending an Urgent Medical Device Correction letter via certified receipt to all affected customers describing the potential issue. Customers were advised to keep the defibrillators in service and to follow the daily Operator's Checklist. All power supplies will be updated. For questions customers were instructed to contact Technical Support at 1-800-442-1142. For questions regarding this recall call 425-867-4000.

Distribution

Worldwide distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MI, MD, MI, NC, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY and the coutnries of ANGUILLA, ARGENTINA, AUSTRALIA, BAHAMAS, BAHRAIN, BARBADOS, BOLIVIA, BRAZIL, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, CYPRUS, DOMINICAN REPUBLIC, EGYPT, EL SALVADOR, GERMANY, GUAM, GUATEMALA, HONG KONG, INDIA, IRAN, ISRAEL, JAMAICA, JORDAN, KUWAIT, LEBANON, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NICARAGUA, PAKISTAN, PANAMA, PARAGUAY, PERU, PUERTO RICO, SAUDI ARABIA, SOUTH AFRICA, SYRIA, TRINIDAD, TRINIDAD AND TOBAGO, TURKEY, UKRAINE, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIRGIN ISLANDS, and YEMEN.

Quantity

53550 units