FDA Recall Terminated

Regimen 28 Day - Glutaraldehyde Sterilizing and Disinfecting Solution

Recall: Z-1721-2019 · Initiated July 13, 2018

Recall

Recall Number
Z-1721-2019
Event Number
82553
Firm
Kerr/Pentron, DBA Kerr Corporation and Pentron Clinical
FEI Number
2024312
Product Code
MED
Status
Terminated
Root Cause
Error in labeling
Initiated
July 13, 2018
Posted
July 5, 2019
Terminated
May 5, 2021
Address
1717 W Collins Ave, Orange, CA, 92867-5422

Description

Regimen 28 Day - Glutaraldehyde Sterilizing and Disinfecting Solution

Reason

Immersion time stated on the bottle is 45 minutes while the product instructions say 90 minutes.

Action

On July 13, 2018, Metrex Research Corp., a subsidiary of Kerr Corporation, sent URGENT NOTICE: Regimen Glutaraldehyde letters to their consignees instructing them to check their inventory and complete the enclosed Field Action Acknowledement and Product Return Form. Distributors were requested to conduct sub-recalls to their consignees.

Distribution

AZ, CA, CO, FL, GA, IL, KY, LA, MA, MN, MO, NJ, NY, OH, PA, SC, TN, TX, UT, VA, and WA.

Quantity

31904