FDA Recall Terminated

OMNIBotics iBlock Cutting Guide, 4146-1000 - Product Usage: The OMNIBotics iBlock Cutting Guide is a stainless-steel surgical instrument used during total knee arthroplasty. Specifically, the iBlock Cutting Guide is used in conjunction with the iBlock Saw Guide with the OMNIBot robotic Motor Unit, and for preparation of the bone for implant implantation. The instrument has been designed to facilitate bone preparation and implantation of the OMNI Apex Knee System.

Recall: Z-1717-2019 · Initiated April 12, 2019

Recall

Recall Number
Z-1717-2019
Event Number
82797
Firm
OMNIlife science Inc.
FEI Number
3008342029
Product Code
OLO
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
April 12, 2019
Terminated
May 14, 2020
Address
480 Paramount Dr, Raynham, MA, 02767-1085

Description

OMNIBotics iBlock Cutting Guide, 4146-1000 - Product Usage: The OMNIBotics iBlock Cutting Guide is a stainless-steel surgical instrument used during total knee arthroplasty. Specifically, the iBlock Cutting Guide is used in conjunction with the iBlock Saw Guide with the OMNIBot robotic Motor Unit, and for preparation of the bone for implant implantation. The instrument has been designed to facilitate bone preparation and implantation of the OMNI Apex Knee System.

Reason

The OMNIBotics iBlock Cutting Guide does not lock onto the iBlock Saw Guide as intended and may loosen during use.

Action

On April 12, 2019, the firm notified customers of the recall through phone, email, and federal mailing titled URGENT MEDICAL DEVICE ADVISORY NOTICE. Customers were informed that the locking mechanism on the iBlock Cutting Guide may loosen during use. Customers were asked to: -Immediately quarantine lot MI1218 -Return product to the firm -Acknowledge receipt of the notice be completing and returning the Receipt form by fax or email. The firm's investigation is ongoing. For further information contact Christina Rovaldi, Sr. Regulatory Affairs Specialist at 774- 226-1847 or 800-448-6664. Company contact hours are Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time.

Distribution

US Nationwide distribution in the states of OK, SC, and WI.

Quantity

9