FDA Recall Terminated

Generation 4 electronic liquid gauges used on the Caire Liberator, Stroller, Sprint, and Hi-Flow Portable oxygen unit.

Recall: Z-1716-2008 · Initiated October 23, 2007

Recall

Recall Number
Z-1716-2008
Event Number
47468
Firm
Caire, Inc.
FEI Number
3004972304
Product Code
BYJ
Status
Terminated
Root Cause
Device Design
Initiated
October 23, 2007
Posted
August 26, 2008
Terminated
December 17, 2009
Address
1800 Sandy Plains Pkwy, Ste 316, Marietta, GA, 30066-6363

Description

Generation 4 electronic liquid gauges used on the Caire Liberator, Stroller, Sprint, and Hi-Flow Portable oxygen unit.

Reason

Moisture ingress into the electronic liquid level indicator may cause inaccurate liquid oxygen level readings. The oxygen could run out without the patient knowledge.

Action

A Technical Service Bulletin was sent to consignees on/about October 23, 2007 to notify customers of the potential issue. Technical support and information for ordering replacement Generation 4 liquid gauges were offered. For technical concerns or if you have not received the notification from the firm, contact U.S. Technical Service at 1-800-482-2473 or European Technical Service at +44 (0) 797 1260077.

Distribution

U.S. Nationwide and the following foreign countries: Argentina, Austria, Brazil, Canada, China, Chile, Central America/Caribbean, Colombia, Costa Rica, Europe, Austria, Belgium, Denmark, France, Germany, Greece, Hungary, Irish Republic, Italy, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland, United Kingdom, Mexico, South America, Peru and Uruguay.

Quantity

3470