FDA Recall Terminated

EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 C) semi-critical endoscopes.

Recall: Z-1714-2015 · Initiated May 11, 2015

Recall

Recall Number
Z-1714-2015
Event Number
71209
Firm
Advanced Sterilization Products
FEI Number
3003662624
Product Code
FEB
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
May 11, 2015
Posted
June 5, 2015
Terminated
July 6, 2016
Address
33 Technology Dr, Irvine, CA, 92618-2346

Description

EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 C) semi-critical endoscopes.

Reason

Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.

Action

A customer notification letter dated 5/7/15 will be sent to all customers on 5/13/15 to inform them that ASP is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels and subsequently pass the cycle without cancellation if there is a disconnected scope. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the attached Business Reply Card (double-sided) via mail or by fax to Stericycle: 888-879-8144. Customers with questions regarding the business reply card are instructed to contact Stericycle at 888-345-4671 and reference event #7781.

Distribution

Worldwide Distribution-US (nationwide) and the countries of Australia, Singapore, Canada, and Thailand.

Quantity

825 units