FDA Recall Terminated

Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting.

Recall: Z-1714-2014 · Initiated May 12, 2014

Recall

Recall Number
Z-1714-2014
Event Number
68363
Firm
Toshiba American Medical Systems Inc
FEI Number
2020563
Product Code
MQB
Status
Terminated
Root Cause
Other
Initiated
May 12, 2014
Posted
June 4, 2014
Terminated
October 2, 2014
Address
2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92780-7047

Description

Toshiba Infinix X-Ray Interventional System INFX-8000V/UU. An X-Ray system that is capable of radiographic and fluoroscopic studies used in an interventional setting.

Reason

Although the Flat Panel Detector (FPD) touch sensor is activated during use, the error message is displayed and the audible signal continues sounding during the table movement of longitudinal tilt, the movement of longitudinal tilt does not stop.

Action

Toshiba America Medical Systems, Inc. initiated this recall by sending recall notifications via USPS return receipt mail. The recall notification letter, dated May 12, 2014, titled "URGENT: MEDICAL DEVICE CORRECTION", informed customers of the following: device information, serial number, SID, reason for reason, purpose of the notification, corrective action, request to customers, contact information, and customer reply form. Customers with questions regarding this letter were instructed to call (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968. For questions regarding this recall call 714-730-5000.

Distribution

Nationwide Distribution including CA and IN

Quantity

2