FDA Recall Terminated

ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, ALTP5P, ALTP_c, ASTP5P and ASTP_c

Recall: Z-1712-2017 · Initiated February 23, 2017

Recall

Recall Number
Z-1712-2017
Event Number
76695
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
JIS
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
February 23, 2017
Posted
March 15, 2017
Terminated
May 9, 2018
Address
500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466

Description

ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, ALTP5P, ALTP_c, ASTP5P and ASTP_c

Reason

Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry Enzyme 2 Calibrator lots 6JD079 and 6MD042.

Action

Siemens mailed an Urgent Medical Device Correction Notice to affected customers on 02/21/2017. The notice informed calibrator customers of the reason for the correction, risk to health, and actions to be taken by customers who are in the process of converting from the fixed factor to the enzyme calibrated assay and information for customers who had already converted, until the investigation is complete.

Distribution

Worldwide Distribution - United States Nationwide: Denmark Sweden Norway Great Britain Belgium Netherlands Spain Israel Switzerland France Portugal Greece Germany Slovenia Czech Republic Hong Kong Colombia Korea Australia

Quantity

314 calibrators