FDA Recall Open, Classified

Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A

Recall: Z-1709-2026 · Initiated February 16, 2026

Recall

Recall Number
Z-1709-2026
Event Number
98530
FEI Number
1417592
Product Code
OGR
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 16, 2026
Posted
April 1, 2026
Address
3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753

Description

Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A

Reason

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

Action

Firm notified customers on 2/16/26. Consignees were provided with a copy of Olympus's notification, which includes reminders to follow the IFU. According to the firm, the component can continue to be utilized by the customer, along with the remainder of the kit.

Distribution

US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.

Quantity

14,379 kits total