Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Recall
- Recall Number
- Z-1709-2022
- Event Number
- 90710
- Firm
- Cook Incorporated
- FEI Number
- 1820334
- Product Code
- DQX
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- July 19, 2022
- Address
- 750 N Daniels Way, Bloomington, IN, 47404-9120
Description
Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Cook notified consignees by Urgent Medical Device Recall Letter on July 19, 2022. Letter states reason for recall, health risk and action to take: Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected].
Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA
1997 units