FDA Recall
Terminated
GE OEC 9600 C-arm Fluoroscopy System, GE Healthcare, Surgery, Salt Lake City, Utah.
Recall: Z-1709-2008
·
Initiated July 2, 2008
Recall
- Recall Number
- Z-1709-2008
- Event Number
- 49483
- Firm
- OEC Medical Systems, Inc
- FEI Number
- 1720753
- Product Code
- JAA
- Status
- Terminated
- Root Cause
- Equipment maintenance
- Initiated
- July 2, 2008
- Posted
- September 20, 2008
- Terminated
- January 20, 2009
- Address
- 384 Wright Brothers Dr, Salt Lake City, UT, 84116-2862
Description
GE OEC 9600 C-arm Fluoroscopy System, GE Healthcare, Surgery, Salt Lake City, Utah.
Reason
Secondary collimator filters may be missing on certain x-ray units. It was discovered that the secondary collimator filter on some OEC 9600 C-arms was omitted when a collimator assembly was replaced. This would reduce the filtration and therefore the Half-Value-Layer (HVL) of the X-ray beam, and may result in non-compliance with respect to Maximum Entrance Exposure Rate (EER) unless the maximum EE
Action
Domestic customer notification letters were sent via certified mail on 07/02/2008. They were advised to have the Maximum EER assessed and to discontinue use if the maximum limits were exceeded.
Distribution
Nationwide Distribution including states of CA, FL, MS, and TX.
Quantity
660 units