FDA Recall Terminated

Toshiba America Medical Systems (TAMS) Aquilion 64 CT system, computed tomography x-ray system.

Recall: Z-1708-2008 · Initiated October 1, 2007

Recall

Recall Number
Z-1708-2008
Event Number
49541
Firm
Toshiba American Med Sys Inc
FEI Number
2020563
Product Code
JAK
Status
Terminated
Root Cause
Device Design
Initiated
October 1, 2007
Posted
September 29, 2008
Terminated
July 16, 2012
Address
2441 Michelle Dr, Tustin, CA, 92780-7047

Description

Toshiba America Medical Systems (TAMS) Aquilion 64 CT system, computed tomography x-ray system.

Reason

Excessive radiation: Intermittent shifting of CT numbers causes the actual dose to the patient to be greater than expected, when Sure Exposure option is used.

Action

Toshiba sent a letter to customers in July 2007, and sent field service representatives to update the Data Aquisition System in September/October 2007.

Distribution

Worldwide distribution, including USA and Japan.

Quantity

1203 total