FDA Recall Terminated

ExacTrac 5.5, an Image Processing System used for patient positioning for radiation therapy or radiosurgery.

Recall: Z-1707-2014 · Initiated April 15, 2014

Recall

Recall Number
Z-1707-2014
Event Number
67926
Firm
Brainlab AG Kapellenstrasse 12 Feldkirchen Germany
FEI Number
3002619595
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
April 15, 2014
Posted
June 3, 2014
Terminated
April 24, 2017

Description

ExacTrac 5.5, an Image Processing System used for patient positioning for radiation therapy or radiosurgery.

Reason

When using a specific plan update workflow in combination with a non-Brainlab treatment planning system (TPS), there is a possibility in ExacTrac 5.5.5 to generate an unintended blend of two initially separate plans of the same patient.

Action

Brainlab sent a "FIELD SAFETY NOTICE / PRODUCT NOTIFICATION" letter dated March 13, 2014 was sent to all affected customers. The letter identifies the affected product, problem, and actions to be taken. The letter includes specific instructions to users that are intended to reduce the likelihood of similar events occurring when using ExacTrac software version 5.5.5. The letter further states that Brainlab will be providing a software update to all customers. Brainlab will contact all customers to arrange for the software update within the following 6 months. Customers with any questions about this issue can contact Brainlab at 800-597-5911 or via e-mail at [email protected].

Distribution

Worldwide Distribution - US Nationwide in the states of AZ, AR, CA, FL, KY, MA, NE, NY, TN, TX and the countries of Argentina, Australia, Brazil, Finland, India, Japan, Netherlands, Oman, Russia, and Taiwan

Quantity

40 systems (worldwide)