FDA Recall
Terminated
Toshiba America Medical Systems (TAMS) Aquilion 32 CT system, computed tomography x-ray system.
Recall: Z-1707-2008
·
Initiated October 1, 2007
Recall
- Recall Number
- Z-1707-2008
- Event Number
- 49541
- Firm
- Toshiba American Med Sys Inc
- FEI Number
- 2020563
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- October 1, 2007
- Posted
- September 29, 2008
- Terminated
- July 16, 2012
- Address
- 2441 Michelle Dr, Tustin, CA, 92780-7047
Description
Toshiba America Medical Systems (TAMS) Aquilion 32 CT system, computed tomography x-ray system.
Reason
Excessive radiation: Intermittent shifting of CT numbers causes the actual dose to the patient to be greater than expected, when Sure Exposure option is used.
Action
Toshiba sent a letter to customers in July 2007, and sent field service representatives to update the Data Aquisition System in September/October 2007.
Distribution
Worldwide distribution, including USA and Japan.
Quantity
1203 total