FDA Recall Terminated

Mobile Diagnostic X-Ray System with Digital Panel

Recall: Z-1706-2017 · Initiated November 28, 2012

Recall

Recall Number
Z-1706-2017
Event Number
64902
Firm
Philips Healthcare Inc.
FEI Number
1218950
Product Code
IZL
Status
Terminated
Root Cause
Software design
Initiated
November 28, 2012
Terminated
May 31, 2018
Address
3000 Minuteman Road, Andover, MA, 01810

Description

Mobile Diagnostic X-Ray System with Digital Panel

Reason

When the operator moves an image to another view then the electronic marker might be placed at the wrong anatomical position in the resulting image.

Action

Philips Medicals Systems distributed an "URGENT Field Safety Notice" alerting customers of the defect on November 28, 2012. Customers are requested to manually rotate images by using the rotation tool.Customers are also recommended to check whether the automatic marker is set correctly, especially after a manual image rotation is performed. The marker can be corrected manually with the annotation tool.

Distribution

Nationwide (AR, AZ, CA, DE, FL, GA, IA, IL, IN, KS, MD, MI, MN, NC, NM, NY, OH, OR, PA, SC, TX, VT, WI)

Quantity

220 systems