FDA Recall
Terminated
Mobile Diagnostic X-Ray System with Digital Panel
Recall: Z-1706-2017
·
Initiated November 28, 2012
Recall
- Recall Number
- Z-1706-2017
- Event Number
- 64902
- Firm
- Philips Healthcare Inc.
- FEI Number
- 1218950
- Product Code
- IZL
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 28, 2012
- Terminated
- May 31, 2018
- Address
- 3000 Minuteman Road, Andover, MA, 01810
Description
Mobile Diagnostic X-Ray System with Digital Panel
Reason
When the operator moves an image to another view then the electronic marker might be placed at the wrong anatomical position in the resulting image.
Action
Philips Medicals Systems distributed an "URGENT Field Safety Notice" alerting customers of the defect on November 28, 2012. Customers are requested to manually rotate images by using the rotation tool.Customers are also recommended to check whether the automatic marker is set correctly, especially after a manual image rotation is performed. The marker can be corrected manually with the annotation tool.
Distribution
Nationwide (AR, AZ, CA, DE, FL, GA, IA, IL, IN, KS, MD, MI, MN, NC, NM, NY, OH, OR, PA, SC, TX, VT, WI)
Quantity
220 systems