Airway Exam Kit, DYKE1796
Recall
- Recall Number
- Z-1705-2026
- Event Number
- 98530
- FEI Number
- 1417592
- Product Code
- CAI
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- February 16, 2026
- Posted
- April 1, 2026
- Address
- 3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753
Description
Airway Exam Kit, DYKE1796
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Firm notified customers on 2/16/26. Consignees were provided with a copy of Olympus's notification, which includes reminders to follow the IFU. According to the firm, the component can continue to be utilized by the customer, along with the remainder of the kit.
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
14,379 kits total