FDA Recall Terminated

AngioDynamics Solero Microwave Tissue Ablation (MTA) System: (1)Solero Applicator 29cm PG Catalog Number:700106003 UPN: H7877001060030 (2)Solero Applicator 29cm PG US Catalog Number:700106003US UPN:H787700106003US0 Product Usage: The Solero Microwave Tissue Ablation (MTA) System and accessories are indicated for the ablation of soft tissue during open procedures. The Solero MTA System is not intended for cardiac use.

Recall: Z-1704-2019 · Initiated April 17, 2019

Recall

Recall Number
Z-1704-2019
Event Number
82743
Firm
Angiodynamics Inc. (Navilyst Medical Inc.)
FEI Number
3017892510
Product Code
NEY
Status
Terminated
Root Cause
Device Design
Initiated
April 17, 2019
Terminated
June 9, 2020
Address
10 Glens Falls, Tech Park Glens Falls, NY, 12801-3864

Description

AngioDynamics Solero Microwave Tissue Ablation (MTA) System: (1)Solero Applicator 29cm PG Catalog Number:700106003 UPN: H7877001060030 (2)Solero Applicator 29cm PG US Catalog Number:700106003US UPN:H787700106003US0 Product Usage: The Solero Microwave Tissue Ablation (MTA) System and accessories are indicated for the ablation of soft tissue during open procedures. The Solero MTA System is not intended for cardiac use.

Reason

Updated Warning Statements in the Operator's Manual/Directions for Use and Software Updates to reduce the risk of potential patient harm due to a delay in performing a procedure

Action

AngioDynamics issued on April 17, 2019 AngioDynamics initiated a Field Correction by issuing notifications via Federal Express. Consignees are instructed to replace the existing Solero Generator User Manual with the revised version provided with the Notification. Once the replacement has been made, the consignees are instructed to complete and return the Reply Verification Tracking Form, provided in the notification. AngioDynamics will also install a software update to help reduce the incidence of Error 0001 which can occur during initial start-up. This improvement will be implemented on the next servicing of the Solero Microwave Tissue Ablation System Generator. Questions regarding the corrective action as stated above, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday Friday: Eastern Standard Time).

Distribution

Worldwide distribution - US Nationwide distribution and countries of AE AT AU BE CA CH CL CO CZ DE DK DZ EG ES FR GB GE GR HK IE IL IN IQ IT LB LT MC MT MX NL NZ PL PT SG TH TR TW VN ZA

Quantity

1846