FDA Recall Open, Classified

2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.

Recall: Z-1703-2026 · Initiated February 25, 2026

Recall

Recall Number
Z-1703-2026
Event Number
98461
Firm
Synthes (USA) Products LLC
FEI Number
3008812560
Product Code
HWX
Status
Open, Classified
Root Cause
Process change control
Initiated
February 25, 2026
Posted
March 30, 2026
Address
1301 Goshen Pkwy, West Chester, PA, 19380-5986

Description

2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US.

Reason

A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.

Action

Depuy Synthes initially notified Sales Consultants and Regional Managers on about 02/25/2026, followed by notification of consignees. Consignees were instructed to examine inventory for affected units, quarantine them immediately, and contact Depuy Synthes sales consultant or customer support services to arrange the return/credits for affected units. Additionally, consignees were requested to complete and return the provided Business Response Form as well as notify all applicable personnel within the facility or any consignees if the affected units were further distributed. Consignees were also requested to post the notification in a visible area for awareness.

Distribution

US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.

Quantity

33 units