FDA Recall Terminated

Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.

Recall: Z-1700-2020 · Initiated February 28, 2020

Recall

Recall Number
Z-1700-2020
Event Number
85275
Firm
Abbott
FEI Number
2017865
Product Code
NIK
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
February 28, 2020
Terminated
April 18, 2024
Address
15900 Valley View Ct, Sylmar, CA, 91342-3577

Description

Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.

Reason

Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained.

Action

On 02/28/2020, the frim sent all US consignees an "IMPORTANT MEDICAL DEVICE CORRECTION" letter via USPS three to five-day ground service. The customer notification informed customer that they had recently deployed a Patient Care System (PCS) programmer software that contained an anomaly that may result in specific programming changes not being stored in the device. The Recalling Firm explained that when using the affected programmer software to make certain programming changes, the changes may appear to have been successfully programmed but the device could remain at previously programmed settings. This issue can occur only if changes to tachycardia therapy Tilt or Waveform (monophasic/biphasic) settings are made and therapy energy value are modified from nominal using the new programmer software. The customer notification also informs customer of Patient Management Recommendations: The Recalling Firm has developed an updated programmer software version which will correct this issue. This solution does not present additional risks to patients and no intervention is needed for patients. The Recalling Firm will update the programmer software in the coming days. For any questions contact your representative or Technical Support at 1-800-722-3774 (U.S.).

Distribution

U.S. AK, Al, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. OUS: None - the firm stated that there are no OUS consignees associated to ICDs that were interrogated from January 22, 2020 to February 6, 2020.

Quantity

Software Version 24.6.1 installed into 4733 programmers