FDA Recall Open, Classified

NexSite HD, Hemodialysis Symmetric Tip Catheter for long term use

Recall: Z-1699-2021 · Initiated February 11, 2020

Recall

Recall Number
Z-1699-2021
Event Number
87827
Firm
MARVAO MEDICAL DEVICES LTD. Unit 1 Galway Business Park Dangan Galway Ireland
FEI Number
3008110587
Product Code
MSD
Status
Open, Classified
Root Cause
Labeling Change Control
Initiated
February 11, 2020

Description

NexSite HD, Hemodialysis Symmetric Tip Catheter for long term use

Reason

The action is to ensure consistency between the Directions for Use (DFU) in the field and the DFU on the website.

Action

On 02/10/2020, the firm emailed a "Direction For Use Update" Letter, to inform them that the firm has updated its Directions for Use (LS 122) from revision 01 to revision 02 and stated that the update is to provide clarification of an existing warning following a complaint where the stylet hub detached following removal without a guidewire in place. The updated DFU now includes the following statement, "The stylet is not to be removed without the guidewire in place. To remove the stylet forcibly in such cases could result in device damage". Customer are begin informed that the Recalling Firm will be removing LS 122 revision 01 which they currently have and replace it with LS 122 revision 02. This "rework" (replacement) will be conducted by the Recalling Firm's distributor DirectACCESS. The rework (replacement) will occur by the end of March 2020. A copy of the DFU is provided on the Recalling Firm's website and is also attached to the customer notification email. For questions associated with this correction - contact Recalling Firm at: Telephone number - +353 (0) 91 759 301 email - [email protected] or, DirectACCESS at (855)210-0554 Ext 114

Distribution

U.S.: CO, NC, SC, TX, O.U.S.: Not provided

Quantity

292 devices