Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant. For in vitro diagnostic use.
Recall
- Recall Number
- Z-1698-2014
- Event Number
- 68065
- Firm
- Leica Microsystems, Inc.
- FEI Number
- 3000210133
- Product Code
- NJT
- Status
- Terminated
- Root Cause
- Incorrect or no expiration date
- Initiated
- February 26, 2014
- Posted
- June 2, 2014
- Terminated
- August 14, 2017
- Address
- 1700 Leider Ln, Buffalo Grove, IL, 60089-6622
Description
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant. For in vitro diagnostic use.
Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected through an ongoing stability program. There is a linkage between the age of the product and staining intensity.
Leica Biosystems sent a Medical Device Recall Notification letter dated February 26, 2014 to all direct accounts. The letter advised that clinicians may want to review results obtained from the affected products. The letter also included instructions for customers to: 1) appropriately destroy any unused or partially used NCL-L-GSTpi-438 (lots 6018777, 6023260,6026297, 6026980) reagent or confirm that they are not in stock; 2) complete and return the attached Medical Device Recall Notification Acknowledgement Form to 847-236-3747; and, 3) contact the local Leica representative immediately if replacement is required. Customers with questions or concerns about this recall can contact Leica via e-mail at [email protected]. For questions regarding this recall call 847-405-5413.
Nationwide Distribution including AZ, CA, FL, and NY.
33 units