FDA Recall Open, Classified

Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: N/A

Recall: Z-1696-2026 · Initiated December 9, 2025

Recall

Recall Number
Z-1696-2026
Event Number
98383
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
NAY
Status
Open, Classified
Root Cause
Process control
Initiated
December 9, 2025
Posted
March 30, 2026
Address
1266 Kifer Rd, Sunnyvale, CA, 94086-5304

Description

Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: N/A

Reason

Due to increased complaints for broken/frayed grip cables for reusable instruments.

Action

On 12/29/2025, the firm sent via email an "Urgent: Medical Device Recall" Letter informing customers that intuitive has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci S and Si reusable instruments. Customers are instruct to: 1. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 2. Identify and quarantine any affected product(s). Affected product(s) listed above may be returned by sending an email with quantities and lot numbers to your regional customer service. 3. Ensure to include the FSCA number ISIFA2025-15-R in the return notes. 4. Credit will be provided based on the number of remaining lives. 5. Shared or further distributed these products with other sites, make sure appropriate staff at the site receive and understand this notification so they locate and return their affected product. For questions or concern, contact Clinical Sales Representative or contact Intuitive Customer Service at the numbers listed below: " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or [email protected] " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or [email protected] " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)

Distribution

Worldwide - U.S. Nationwide distribution in the states of CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MO, MS, NC, NJ, NY, OH, PA, TX, VA and WI. The countries of Argentina, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Ecuador, France, Germany, Greece, Guadeloupe, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Pakistan, Philippines, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, South Korea, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.

Quantity

86904