FDA Recall Terminated

Brilliance iCT Computed Tomography X-ray system

Recall: Z-1695-2015 · Initiated April 13, 2015

Recall

Recall Number
Z-1695-2015
Event Number
71231
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Other
Initiated
April 13, 2015
Posted
May 29, 2015
Terminated
October 1, 2015
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

Brilliance iCT Computed Tomography X-ray system

Reason

A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.

Action

Philips sent an Urgent - Field Safety Notice dated April 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips Healthcare is recommending to stop the use of your system until the X-ray tube mounting fasteners are replaced. In addition, a Philips Field Service Engineer (FSE) will service your system via a Mandatory Field Change Order (FCO 72800638). A Philips FSE will contact you to schedule this appointment. If you need any further information or support concerning this issue, please contact your local Philips representative or your local Philips Healthcare office. This notice has been submitted to the appropriate Regulatory Agency. Philips apologizes for any inconveniences this problem has caused your organization. For further questions please call (440) 483-7600.

Distribution

Internationally to the countries of China and Norway.

Quantity

2