FDA Recall
Terminated
Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120 volt configuration only.
Recall: Z-1693-2009
·
Initiated July 29, 2009
Recall
- Recall Number
- Z-1693-2009
- Event Number
- 52835
- Firm
- Draeger Medical Systems, Inc.
- FEI Number
- 2510954
- Product Code
- FMT
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- July 29, 2009
- Posted
- August 26, 2009
- Terminated
- October 13, 2010
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042
Description
Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120 volt configuration only.
Reason
Spark, flame and smoke: A mis-positioned crimp on heater element lead wire resulted in flame and smoke aerating from the top of the warmer assembly.
Action
Draeger Medical issued an Urgent Medical Device Recall letter via certified mail on 8/7/09. The letter explained the problem, the need to conduct an inspection of the infant warmer (instructions included), and request replacement elements.
Distribution
Worldwide distribution: USA, Canada, Saudi Arabia, Bermuda, Virgin Islands, Mexico, Panama, Colombia, and Costa Rica.
Quantity
2915 elements