FDA Recall Terminated

Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120 volt configuration only.

Recall: Z-1693-2009 · Initiated July 29, 2009

Recall

Recall Number
Z-1693-2009
Event Number
52835
Firm
Draeger Medical Systems, Inc.
FEI Number
2510954
Product Code
FMT
Status
Terminated
Root Cause
Process control
Initiated
July 29, 2009
Posted
August 26, 2009
Terminated
October 13, 2010
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120 volt configuration only.

Reason

Spark, flame and smoke: A mis-positioned crimp on heater element lead wire resulted in flame and smoke aerating from the top of the warmer assembly.

Action

Draeger Medical issued an Urgent Medical Device Recall letter via certified mail on 8/7/09. The letter explained the problem, the need to conduct an inspection of the infant warmer (instructions included), and request replacement elements.

Distribution

Worldwide distribution: USA, Canada, Saudi Arabia, Bermuda, Virgin Islands, Mexico, Panama, Colombia, and Costa Rica.

Quantity

2915 elements