FDA Recall Terminated

iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538

Recall: Z-1693-2008 · Initiated April 11, 2000

Recall

Recall Number
Z-1693-2008
Event Number
47759
Firm
Boston Scientific Corporation
FEI Number
2939204
Product Code
IYO
Status
Terminated
Root Cause
Process control
Initiated
April 11, 2000
Posted
September 23, 2008
Terminated
April 6, 2009
Address
47900 Bayside Pkwy, Fremont, CA, 94538-6515

Description

iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538

Reason

Improperly terminated wires on a component of the iLab Acquisition processor power supply could cause the processor to lock-up or stop, and prolong the patients procedure.

Action

On April 4 and 10, 2008, customer notifcations were sent to consignees. Field corrections are estimated to begin on April 25, 2008.

Distribution

Product was distributed to a total of 81 consignees in the USA, Hong Kong, Great Britain, Austria and France.

Quantity

120 units