FDA Recall
Terminated
iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538
Recall: Z-1693-2008
·
Initiated April 11, 2000
Recall
- Recall Number
- Z-1693-2008
- Event Number
- 47759
- Firm
- Boston Scientific Corporation
- FEI Number
- 2939204
- Product Code
- IYO
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- April 11, 2000
- Posted
- September 23, 2008
- Terminated
- April 6, 2009
- Address
- 47900 Bayside Pkwy, Fremont, CA, 94538-6515
Description
iLab Ultrasound Imaging System, Models 120INS and 240INS, manufactured by Boston Scientific Corporation, Fremont, CA 94538
Reason
Improperly terminated wires on a component of the iLab Acquisition processor power supply could cause the processor to lock-up or stop, and prolong the patients procedure.
Action
On April 4 and 10, 2008, customer notifcations were sent to consignees. Field corrections are estimated to begin on April 25, 2008.
Distribution
Product was distributed to a total of 81 consignees in the USA, Hong Kong, Great Britain, Austria and France.
Quantity
120 units