FDA Recall Terminated

NeoBase Succinylacetone Assay Solution, Product Number: 3042-0020; Affected UDI Numbers: (01)06438147295708(17)180531(10)654184 (01)06438147295708(17)180531(10)653414 (01)06438147295708(17)180531(10)653657 (01)06438147295708(17)180630(10)654923 (01)06438147295708(17)180630(10)655680 (01)06438147295708(17)180630(10)655300 (01)06438147295708(17)180630(10)654924 (01)06438147295708(17)180831(10)656025 (01)06438147295708(17)180831(10)657055 (01)06438147295708(17)180831(10)656170 (01)06438147295708(17)180831(10)656320 (01)06438147295708(17)181031(10)657768 (01)06438147295708(17)181031(10)657513 (01)06438147295708(17)181031(10)658006

Recall: Z-1692-2018 · Initiated November 2, 2017

Recall

Recall Number
Z-1692-2018
Event Number
79303
Firm
PerkinElmer Life and Analytical Sciences, Wallac, OY Mustionkatu 6
FEI Number
3002808498
Product Code
NQL
Status
Terminated
Root Cause
Process control
Initiated
November 2, 2017
Terminated
September 17, 2019
Address
P.O. Box 10, Turku Finland

Description

NeoBase Succinylacetone Assay Solution, Product Number: 3042-0020; Affected UDI Numbers: (01)06438147295708(17)180531(10)654184 (01)06438147295708(17)180531(10)653414 (01)06438147295708(17)180531(10)653657 (01)06438147295708(17)180630(10)654923 (01)06438147295708(17)180630(10)655680 (01)06438147295708(17)180630(10)655300 (01)06438147295708(17)180630(10)654924 (01)06438147295708(17)180831(10)656025 (01)06438147295708(17)180831(10)657055 (01)06438147295708(17)180831(10)656170 (01)06438147295708(17)180831(10)656320 (01)06438147295708(17)181031(10)657768 (01)06438147295708(17)181031(10)657513 (01)06438147295708(17)181031(10)658006

Reason

There is a potential for leaking vials for certain lot numbers where, if proper attire is not used as instructed, a transient minor allergic skin reaction may result when handling leaking succinylacetone reagent vials.

Action

Actions to be taken: - Inspect your inventory of the NeoBase Succinylacetone Assay Solution vials for the affected lots. - If you notice leakage or deformed stopper seals, destroy the affected vials according to local instructions and regulations. - Complete the Response Form with the quantity and identity of affected Succinylacetone Assay Solution vials in your inventory. - Return the Response Form to PerkinElmer. - PerkinElmer will replace the quantity of leaking Succinylacetone Assay Solution vials found and destroyed from your inventory

Distribution

Distributed in 20 US states: AL, AR, AZ, CA, CO, FL, GA, IL, IN, KS, KY, LA, MI, MO, NV, OH, OR, SC, TN, TX.

Quantity

5275