GE Innova 3100-IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector). Indicated for use in generating fluoroscopic images of human anatomy for vascular angiography diagnostic and interventional procedures and optionally, rotational imaging procedures. It is also intended for generating fluoroscopic images of human anatomy for cardiology, diagnostic and interventional procedures.
Recall
- Recall Number
- Z-1689-2009
- Event Number
- 53260
- Firm
- Ge Healthcare
- FEI Number
- 3005860720
- Product Code
- MQB
- Status
- Terminated
- Root Cause
- Radiation Control for Health and Safety Act
- Initiated
- October 9, 2008
- Posted
- September 24, 2009
- Terminated
- January 7, 2012
- Address
- 9900 W Innovation Dr, Wauwatosa, WI, 53226-4856
Description
GE Innova 3100-IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector). Indicated for use in generating fluoroscopic images of human anatomy for vascular angiography diagnostic and interventional procedures and optionally, rotational imaging procedures. It is also intended for generating fluoroscopic images of human anatomy for cardiology, diagnostic and interventional procedures.
GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE Healthcare found the AKR calibration procedure does not exactly reflect the worst-case clinical usage of the product. This may result, when using large Equival
Consignees were sent a GE Healthcare "Important electronic Product Radiation Warning" letter dated November 3, 2008. The letter was addressed to Healthcare Administrator/Risk Managers, Managers of Radiology/Cardiology, Radiologists/Cardiologists. The letter described the Defect Description and Related Hazards, Affected Product Details, Safety Instructions, Product Correction and Contact Information. For further information, contact the GE Healthcare Technical Support Center at 1-800-437-1171.
Nationwide Distribution -- United States including Puerto Rico and Guam.
929 units total