FDA Recall Terminated

VITROS CoV2 Ag calibrator (Product Code 6199950) - Product Usage: is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) and anterior nasal swab specimens collected in CDCs formulation of VTM.

Recall: Z-1688-2021 · Initiated May 3, 2021

Recall

Recall Number
Z-1688-2021
Event Number
87888
Firm
Ortho-Clinical Diagnostics, INc.
FEI Number
1000305840
Product Code
QKP
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 3, 2021
Terminated
October 24, 2023
Address
513 Technology Blvd, Rochester, NY, 14626-3601

Description

VITROS CoV2 Ag calibrator (Product Code 6199950) - Product Usage: is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) and anterior nasal swab specimens collected in CDCs formulation of VTM.

Reason

Ortho investigated and confirmed the potential for unexpected failed calibration, and condition codes associated with the calibration, when using VITROS SARS-CoV-2 Ag Reagent Pack along with VITROS Immunodiagnostic Products SARS-CoV-2 Ag Calibrator, Lot 20.

Action

On May 4, 2021, Ortho Clinical Diagnostics issued an Urgent Medical Device Correction notice to customers via letter notifying them of calibration failures and the associated condition code for specified calibrator lots. Customers received the following instructions. " Discontinue using VITROS SARS-CoV-2 Antigen Reagent/Calibrator Lot 0064, discard any remaining inventory and switch to an alternate lot. " Ortho will credit and/or replace all remaining inventory of VITROS SARS-CoV-2 Antigen Reagent/Calibrator Lot 0064 as indicated on the Confirmation of Receipt form. " Complete the enclosed Confirmation of Receipt form no later than MM/DD/YY, even if your facility no longer has the affected product in inventory. " Please forward this notification if the product was distributed outside of your facility.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AR, CA, IA, IL, LA, MO, MS, TX, WI, and the countries of Colombia, India, Japan, Italy, Spain, Portugal, France.

Quantity

74 units